spacer
spacerspacer
HomeCheckout
0 £0.00
Actim Partus Home Button
Why Actim Partus?
Actim Partus Efficiency
Test Procedure
FAQ's
Downloads & Reference
Partus Links
Buy Online button
spacer
Products Button
CardioChek PA
Medi-Test
Clover A1
Biocard Celiac Test
Check-Mate
FastCheck Allergy Tests
Stat-Site
Actim Flu
Actim Prom
Actim Fecal Blood
spacer
Products by sector button
Cardiovascular
Diabetes
Female Health
Gastro-intestinal
Haematology
Urine Analysis
spacer
Site Switcher Button
spacer
About BHR
Contact BHR
protx_logo
Home > Actim Partus > Actim Partus Test Procedure

Actim Partus Test Procedure

actim prom test strip detail
test procedure 1gutter spacerA cervical secretion sample is taken with a sterile Dacron swab from the cervix by leaving the Dacron swab in the cervix for ten to fifteen seconds. This allows the swab to absorb the cervical secretion specimen.
test procedure 2gutter spacerThe Dacron swab is then placed in the extraction solution provided and swirled around vigorously for ten seconds.
test procedure 3gutter spacerAfter extraction, dip the yellow area of the dipstick into the extraction solution and hold it there until the liquid front reaches the result area.

Then simply remove the dipstick from the solution and place it in a horizontal position.

test procedure 4gutter spacerA positive test result can be seen as soon as two blue lines - a control line and a positive line - appear in the result area.

If, after five minutes, only the control line has appeared, the test result is negative.