Actim Prom Test Procedure

actim prom test strip detail
test procedure 1gutter spacerA vaginal fluid sample is taken with a Dacron swab from the posterior fornix of the vagina - or from the cervix if no fluid is visible - by leaving the Dacron swab in the vagina for ten to fifteen seconds. This allows the swab to absorb sufficient amount of specimen.
test procedure 2gutter spacerThe Dacron swab is then placed in the extraction solution provided and swirled around vigorously for ten seconds.
test procedure 3gutter spacerAfter extraction, dip the yellow area of the dipstick into the extraction solution and hold it there until the liquid front reaches the result area. Then simply remove the dipstick from the solution and place it in a horizontal position.
test procedure 4gutter spacerA positive test result can be seen as soon as two blue lines - a control line and a positive line - appear in the result area.

If, after five minutes, only the control line has appeared, the test result is negative.